M561-00-880 Issue C
Press Stop
button PDT displays:
Pump State
Local Control LED
Stopped
On
Running
On
Stopped or Running
Off
A1.3
Warning / Alarm display and acknowledgement
Each new warning / alarm is displayed when it occurs, overwriting any text already present unless another
unacknowledged warning / alarm is currently being displayed. The corresponding Warning / Alarm LED flashes to
indicate a new warning / alarm.
Pressing ENTER acknowledges the warning / alarm currently displayed and the Warning / Alarm LED stops flashing.
If available the display will show the suggested action, press ENTER again to clear.
If there is another new warning or alarm the Warning / Alarm LED will continue to flash and this is then displayed,
otherwise the display will revert to the original text from before the alarm(s) / warning(s) occurred.
If there are warnings / alarms still present, but they are all acknowledged, then the corresponding LED remains
continuously illuminated. The text indicating acknowledged warning / alarm conditions still present may be viewed
in the Status menu, see below. Some Alarms such as 1.01 STOP ACTIVATED remain present until the pump is manually
started from the PDT or front panel controls.
Once all warning/alarm conditions have gone away, then the corresponding LED is extinguished.
To avoid a build up of out of date Warnings they are automatically acknowledged after 36 hours.
A1.4
Menus
There are three menu buttons Normal, Status and Setup described below.
NR in the data part of a parameter indicates No Reading.
NP in the data part of a parameter indicates a parameter that is Not Present.
A1.4.1

Normal menu

This menu is displayed when the PDT is first plugged into the pump, or accessed by pressing the Normal button. Up
to 4 parameters are displayed, scroll by pressing up/down keys.
Description
Serial Number
Control Holder
Dry Pump current
Booster current
Page 58
Table A2 - Pump stop and control
PDT display
PUMP Stopped Press
CANCEL
STOP MENU Fast
Shutdown
No PDT Control Press
CANCEL
Table A3 - Normal menu
Typical Display
S/N 1234567
NONE IN CONTROL
DP CURRENT 1.1A
MB CURRENT 1.1A
Edwards and the Edwards logo are trademarks of Edwards Limited.
Operator
Pump response
Press CANCEL
No change (stopped)
If Auto Shutdown is
Pump Stops
required press down
arrow. Press ENTER
to confirm
Press CANCEL
No change
© Edwards Limited 2012. All rights reserved.