Extraction Rates Required By System Variant - Edwards iXH100 Instruction Manual

Dry pumping systems
Table of Contents

Advertisement

M561-00-880 Issue C
Extraction rate required at top of enclosure
System
3
(m
/h)
iXH100
180
iXH200H
180
iXH450H
310
iXH500H
310
iXH610
180
iXH1210/H
180
iXH1220H/T
180
iXH1820/H/T
180
iXH3030/T
310
iXH3045H
310
iXH3050H
310
iXH4545HT
310
iXH4550HT
310
iXH6045H/T
310
iXH6050H/T
310
iXL250Q
180
iXL500Q
310
iXL500R
310
iXL750Q
310
iXL750R
310
Page 22
Table 18 - Extraction rates required by system variant
Edwards and the Edwards logo are trademarks of Edwards Limited.
Extraction rate required at exhaust cover
3
(m
/h)
0
0
60
60
0
0
0
0
60
60
60
60
60
60
60
60
60
60
60
310
© Edwards Limited 2012. All rights reserved.