Wire Assembly According To Vde 0295 - Edwards iXH100 Instruction Manual

Dry pumping systems
Table of Contents

Advertisement

Description
Accessory interface
Analogue measurement for water flow meter
Pin 1 - input, Pin 5 common
Active accessory module
Pin 3 - RS485 +, Pin 10 - RS485 –
Pump running status contacts
Pin 6 - Dry pump (normally open)
Pin 14 - Mechanical booster (normally open)
Pin 15 – common
Gate valve
Pin 4 - Gate valve drive transistor (open collector)
Gate valve position sense
Pin 7 - 'Closed', Pin 8 - 'Open'
Power Supplies
Pin 12 - Accessory 24 V supply
Pin 13 - Accessory 24 V supply
Pin 5 - 0 V supply common
*
The System interface, the EMS interface and the Accessory interface have a combined current rating of 0.75 A.
This supply will be disconnected in the event of an emergency stop.
Wire Size (mm²)
Table 9 - Electrical connections (continued)
Mating Connector description / external supply rating
Table 10 - Wire assembly according to VDE 0295
6
10
16
25
35
50
15-way D socket
30 V a.c. 1 A, 60 V d.c. 0.5 A
Fine stranded wires VDE 0295 class 5
Internal supply rating
24 V d.c. 0.75 A
24 V d.c. 0.2 A
84 x 0.30
80 x 0.40
128 x 0.40
200 x 0.40
280 x 0.40
400 x 0.40
*

Hide quick links:

Advertisement

Table of Contents
loading