Edwards iXH100 Instruction Manual page 79

Dry pumping systems
Table of Contents

Advertisement

Warning Message
'Action' Message on
on PDT
PDT
Warning 70.13
-
Sensor missing
Warning 71.13
See manual
AC Sup Missing
Warning 84.11
See manual
Exh Temp High
Warning 84.13
-
Sensor missing
Warning 152.01
Check Gate Valve
Valve Not Shut
Warning 153.01
Check Gate Valve
Valve Not Open
Warning 176.01
xxxx yyyy
MB INV xxxx yyyy
aaaaaaaaaaaaaaaa
Warning 176.13
See manual
No MB Inv Comms
Warning 186.01
xxxx yyyy
DP INV xxxx yyyy
aaaaaaaaaaaaaaaa
Warning 186.13
See manual
No DP Inv Comms
Warning 196.01
Diag aaaa bbbb
DP Inv xxxx yyyy
/zzzzzzzzzzzzzzz
Warning 196.01
See manual
DP Not Running
Warning 196.13
See manual
No DP Inv Comms
Warning 314.11
Stop Pump/Contact
Speed Too Low
Service
Notes:
1. Some of these warnings apply to pump sensors that are only present on certain system variants.
2. Some of these warnings apply to accessories and may only be encountered if the particular accessory is
fitted.
© Edwards Limited 2012. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Table A15 - Warnings (continued)
Possible Cause
The dry pump temperature sensor may
have become disconnected or failed.
The pump controller cannot communicate
with the accessory module
The exhaust pipe temperature is too high
The exhaust pipe temperature sensor may
have become disconnected or failed.
The foreline gate valve has failed to close
The foreline gate valve has failed to open
Booster Inverter has raised a warning code. Refer to
The pump controller cannot communicate
with the booster inverter.
Dry pump Inverter has raised a warning
code.
The pump controller cannot communicate
with the dry pump inverter.
DP 2nd Inverter has raised warning code.
DP 2nd Inverter will not start up when
requested.
The pump controller cannot communicate
with the dry pump 2nd inverter.
The pump rotational speed is at least 30%
less than demand for more than 2 minutes.
M561-00-880 Issue C
Action
- Check that the sensor is
fitted and is correctly
connected.
- Check that the accessory
module is correctly
connected to the pump and
then cycle the power to the
pump.
- If this warning persists,
contact Edwards.
Contact Edwards
Contact Edwards
Check the wiring and air
supply to the Gate valve.
Note:
This warning will
only clear when the
valve has
successfully been
opened and closed.
Check the wiring and air
supply to the Gate valve.
Note:
This warning will
only clear when the
valve has
successfully been
closed and opened.
A2.3
for more
information.
Check the wiring between
the pump controller and the
booster inverter.
Refer to
A2.3
for more
information.
Check the wiring between
the pump controller and the
dry pump inverter.
Refer to
A2.3
for more
information.
Check Inverter Fault History.
Check the wiring between
the pump controller and the
dry pump inverter.
Stop the pump and perform
mechanical check of
rotation and beearings.
Page 67