2.5

Electrical data

Characteristics
iXH100
Dry pump motor rating
4.5
Mechanical booster
-
motor rating
Current rating
21
(200 - 230 V systems)
Current rating
11
(380 - 460 V systems)
Recommended branch
25
circuit fuse UL
(200 - 230 V systems)
Recommended branch
25
circuit fuse IEC
(200 - 230 V systems)
Recommended branch
15
circuit fuse UL
(380 - 460 V systems)
Recommended branch
15
circuit fuse IEC
(380 - 460 V systems)
*
Min cable size for 200 -
6 (8)
230 V systems (or
corresponding AWG
size)
*
Min cable size for 380 -
6 (8)
460 V systems (or
corresponding AWG
size)
®
Mains connector
Han
K 4/4
*
The minimum geometric wire gauge for Han
Table 7 - Electrical data iXH100 to iXH1820/H/T
iXH200H
iXH450H
iXH500H
7.5
11
11
-
-
-
38
44
44
20
23
23
45
55
55
40
45
45
25
30
30
20
25
25
10 (6)
10 (6)
10 (6)
*
*
*
6 (8)
6 (8)
6 (8)
®
®
®
Han
K 4/4 Han
K 4/4
Han
K 4/4
2
®
K 4/4 is 6mm
and minimum AWG size is 8 AWG
iXH610
iXH1210
iXH1210H
4.5
4.5
7.5
1.9
4.5
4.5
28
28
38
14
14
20
35
35
45
30
30
40
20
20
25
15
15
20
*
*
6 (8)
6 (8)
10 (6)
*
*
*
6 (8)
6 (8)
6 (8)
®
®
®
Han
K 4/4
Han
K 4/4
Han
K 4/4
iXH1220H/T iXH1820/H/T
Units
7.5
7.5
kW
4.5
4.5
kW
38
38
A
20
20
A
45
45
A
40
40
A
25
25
A
20
20
A
2
10 (6)
10 (6)
mm
(AWG)
*
*
2
6 (8)
6 (8)
mm
(AWG)
®
®
Han
K 4/4
Han
K 4/4
-