Appendix A
Error Messages

Displaying Error Messages

Error List

Error
Device-dependent Errors (these errors set Standard Event Status register bit #3)
0
No error
This is the response to the ERR? query when there are no errors.
100
Too many channels
You have specified more channels than are installed in the mainframe.
101
Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
102
Calibration password is incorrect
The calibration password is incorrect.
103
Calibration is inhibited by switch setting
Calibration mode is locked out by the calibration switch.
104
Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
105
Unexpected output current
The measured output current is outside the acceptable range.
106
Zero measurement out of range error
The "zero" measurement value is outside the acceptable range.
107
Programming cal constants out of range
The programmed calibration constant is outside the acceptable range.
108
Measurement cal constants out of range
The measurement calibration constant is outside the acceptable range.
109
Over voltage cal constants out of range
The over voltage calibration constant is outside the acceptable range.
110
Wrong V+I
The instrument was unable to set the correct voltage or current value.
111
Aux vloc cal constants out of range
Calibration constants on the internal auxiliary local ADC are outside the acceptable range.
112
Aux vrem cal constants out of range
Calibration constants on the internal auxiliary remote ADC are outside the acceptable range.
113
Aux imon cal constants out of range
Calibration constants on the internal auxiliary imon ADC are outside the acceptable range.
114
Wrong status
An incorrect status function has been reported.
126
The instrument beeps each time a command syntax or hardware error is
generated. The front-panel ERR annunciator turns on when one or more
errors are currently stored in the SCPI error queue.
Front Panel:
Press the Error key.
If errors appear, use the navigation
keys to scroll though the list.
Errors are cleared as they are read. When all errors have been read, the
ERR annunciator turns off and the error queue is cleared.
SCPI Command:
SYST:ERR?
Model N6705 Service Guide