Calibration
Calibration Procedure – Models N678xA SMU
Enter Calibration mode
Voltage Calibration
62
Front Panel:
Press [Menu], then select Utilities,
Administrative Tools, Administrator Login.
Enter your numeric password in the PIN field
and select Login.
Then select the Calibration menu. Select
Turn On/Off to access the menu items.
High Range Programming and Measurement - All Models
Step 1. Connect the voltage input of the Keysight 3458A to an output channel
(see Cal Setup A).
Step 2. Select the full-scale voltage programming and measurement range. The
value entered must be the maximum voltage of the range that you are
calibrating. This example selects the 20 V full-scale range of output 1.
Front Panel:
From the Calibration menu, select Voltage,
High Range Prog and Meas. Then select
Start.
Step 3. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 4. Measure the output voltage with the DMM and enter the data. This value
may be positive or negative.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 5. Select the second voltage calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should
be displayed.
SCPI Command:
CAL:STAT ON,
<password>
SCPI Command:
CAL:VOLT 20
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?