Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Low Range Voltage Measurement
Step 1. Select the low-voltage measurement range. The value entered must be
the maximum voltage of the range to be calibrated. This example selects
the 5 V measurement range of output 1.
Front Panel:
From the Calibration menu, select
Voltage, Low Range Measurement. Then
select Start.
Step 2. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should
be displayed.
Step 3. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 4. Select the second voltage calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should
be displayed.
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Calibration
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:VOLT:MEAS 5
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
3
55