Enter a Calibration Date
Save and Log out of Calibration Mode
CAUTION
Model N6705 Service Guide
Step 3. Select the first current calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 4. Calculate the shunt current (I=V/R); enter the negative current value.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 5. Select the second current calibration point.
Front Panel:
The second calibration point is automatically
selected "Enter P2 measured data" should be
displayed.
Step 6. Calculate the shunt current (I=V/R); enter the negative current value.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Front Panel:
From the Calibration menu, select
Date. Enter the date in the format
"yyyy/mm/dd". Then select Insert
Date. Then select Close.
Storing calibration constants overwrites the existing ones in non-volatile
memory. If you are not sure you want to permanently store the new
constants, do not Save the data when you exit the calibration mode. The
calibration will then remain unchanged.
Front Panel:
From the Calibration menu, select
Save. Then select Save Calibration for
all Outputs. Select Close.
Go [Back] to the Administrative tools
menu and select Administrator Login.
Select Logout, then Close
Calibration
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
To enter a calibration date:
CAL:DATE "<date>",
SCPI Command:
To save calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
3
(@1)
79