3
Calibration
Enter a Calibration Date
Save and Log out of Calibration Mode
CAUTION
48
Step 6. Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
Enter the calculated value into the data field
and then select Next. When calibration is
complete, select Close.
Front Panel:
From the Calibration menu, select
Date. Enter the date in the format
"yyyy/mm/dd". Then select Insert
Date. Then select Close.
Storing calibration constants overwrites the existing ones in non-volatile
memory. If you are not sure you want to permanently store the new
constants, do not Save the data when you exit the calibration mode. The
calibration will then remain unchanged.
Front Panel:
From the Calibration menu, select
Save. Then select Save Calibration
for all Outputs. Select Close.
Go [Back] to the Administrative
tools menu and select
Administrator Login. Select Logout,
then Close
SCPI Command:
CAL:DATA <data>
SCPI Command:
To enter a calibration date:
CAL:DATE "<date>",
SCPI Command:
To save calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF
Model N6705 Service Guide