Calibration Procedure - Models N6783A-Bda And -Mfg - Keysight N6705 Service Manual

Dc power
Table of Contents

Advertisement

3
Calibration
Calibration Procedure – Models N6783A-BDA and -MFG
Enter Calibration mode
Voltage Calibration
76
Front Panel:
Press [Menu], then select Utilities,
Administrative Tools, Administrator Login.
Enter your numeric password in the PIN field
and select Login.
Then select the Calibration menu. Select
Turn On/Off to access the menu items.
High Range Programming and Measurement
Step 1. Connect the voltage input of the Keysight 3458A DMM to an output
channel (see Cal Setup A).
Step 2. Select the full-scale voltage programming and measurement range. The
value entered must be the maximum voltage of the range that you are
calibrating. The following example selects the 6 V full-scale range of
output 1. Full-scale ranges vary according to model.
Front Panel:
From the Calibration menu, select Voltage,
High Range Prog and Meas. Then select
Start.
Step 3. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 4. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 5. Select the second voltage calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should
be displayed.
SCPI Command:
CAL:STAT ON,
<password>
SCPI Command:
CAL:VOLT 6
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
Model N6705 Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents