Calibration
Downprogrammer Calibration
Enter a Calibration Date
Save and Log out of Calibration Mode
CAUTION
60
Front Panel:
From the Calibration menu, select
Miscellaneous, Peak Current Limit. Then
select Start.
Step 3 Select the first current calibration point.
Front Panel:
The calibration is automatic.
Step 4. Set the electronic load or connect the appropriate resistor for the second
calibration point.
Step 5. Select the second current calibration point.
Front Panel:
Select Next. The calibration is automatic.
Then select Close.
Step 1. Remove all loads from the output. This procedure is automatic and takes
a few seconds.
Front Panel:
From the Calibration menu, select
Miscellaneous, Downprogrammer. Select
Start. Then select Close.
Front Panel:
From the Calibration menu, select
Date. Enter the date in the format
"yyyy/mm/dd". Then select Insert
Date. Then select Close.
Storing calibration constants overwrites the existing ones in non-volatile
memory. If you are not sure you want to permanently store the new
constants, do not Save the data when you exit the calibration mode. The
calibration will then remain unchanged.
SCPI Command:
CAL:CURR:PEAK
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DPRog
(@1)
*OPC?
SCPI Command:
To enter a calibration date:
CAL:DATE "<date>",
Model N6705 Service Guide
(@1)
(@1)