5
Troubleshooting
B
Is this a spec
failure?
NO
Self test
failure?
NO
Is module in
unregulated mode?
NO
Is protection
falsely tripping?
NO
Can't calibrate
one output
NO
Can't calibrate
all modules
NO
C
108
YES
YES
YES
YES
YES
YES
Figure 5-1. General Troubleshooting Sheet 2
1. Calibrate Output Module
2. Replace the Output Module
1. Calibrate Output Module
2. Replace Output Module
1. Calibrate Output Module
2. Replace Output Module
1. Check to ensure that protection condition is not present
2. Replace Output Module
1. Update firmware
2. Check calibration switches on Trigger Assembly
3. Replace Output Module
1. Check jumpers on Trigger Assembly
2. Reset password using jumpers on Trigger Assembly
3. Update firmware
4. Replace PPMC Assembly
Model N6705 Service Guide