Keysight N6705 Service Manual page 30

Dc power
Table of Contents

Advertisement

2
Performance Verification
Test Record Form – Keysight N6754A and N6756A
Keysight N6754A and N6756A
Description
Voltage Programming & Readback
Minimum Voltage Vout
Voltage measured over interface
High Voltage Vout
Voltage measured over interface
CV Load Effect
CV Source Effect
CV Ripple and Noise
peak-to-peak
rms
Transient Response
Voltage at 100 µs
Voltage at 100 µs w/Option
760/761
Current Programming & Readback
Minimum Current Iout
Current measured over interface
High Current Iout
Current measured over interface
CC Load Effect
CC Source Effect
Test Description
Voltage Programming & Readback, Min Voltage
Voltage Programming & Readback, High Voltage
CV Load Effect, Source Effect, Ripple and Noise
Transient Response
Current Programming & Readback, Min Current
Current Programming & Readback, High Current
CC Load Effect, Source Effect
30
Report No _______________
Model
Minimum Specs.
Constant Voltage Tests
Both
0 mV
Vout − 25 mV
Both
Both
59.939 V
Both
Vout − 55 mV
− 2 mV
Both
− 1.2 mV
Both
Both
N/A
Both
N/A
− 90 mV
Both
− 350 mV
Both
Constant Current Tests
Both
+ 8 mA
Iout − 8 mA
Both
N6754A
19.968 A
N6756A
16.071 A
Iout − 28 mA
N6754A
N6756A
Iout − 25 mA
− 5 mA
Both
− 2 mA
Both
N6754A
25 mV, 5 A
60 V, 5 A
60 V, 5 A
15 V, from 10 A to 20 A
20 mA, 15 V
20 A, 15 V
20 A, 15 V
Date __________________
Results
Maximum Specs.
__________
+ 50 mV
__________
Vout + 25 mV
__________
60.061 V
__________
Vout + 55 mV
__________
+ 2 mV
__________
+ 1.2 mV
__________
+ 6 mV
__________
+ 1 mV
__________
+ 90 mV
__________
+ 350 mV
__________
+ 32 mA
__________
Iout + 8 mA
__________
20.032 A
17.029 A
__________
Iout + 28 mA
Iout + 25 mA
__________
+ 5 mA
__________
+ 2 mA
Instrument Settings
N6756A
25 mV, 8.33 A
60 V, 8.33 A
60 V, 8.33 A
29 V, from 8.5 A to 17 A
20 mA, 29 V
17 A, 29 V
17 A, 29 V
Model N6705 Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents