Keysight N6705 Service Manual page 63

Dc power
Table of Contents

Advertisement

Model N6705 Service Guide
Step 6. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. If
calibration is complete, select Close.
Steps 7 and 8 only apply to model N6784A.
Step 7. Select the third voltage calibration point for the negative range.
Front Panel:
The third calibration point is automatically
selected. "Enter P3 measured data" should
be displayed.
Step 8. Measure the output voltage with the DMM and enter the data. The value
should be negative.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Low Range Voltage Programming - Models N6781A, N6782A,
N6784A
Step 1. Select the low-voltage programming range. The value entered must be
the maximum voltage of the range to be calibrated. This example selects
the 6 V range of output 1.
Front Panel:
From the Calibration menu, select Voltage,
Low Range Programming. Then select Start.
Step 2. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should
be displayed.
Step 3. Measure the output voltage with the DMM and enter the data. This value
may be positive or negative.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Calibration
3
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P3
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:VOLT 6
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
63