Model N6705 Service Guide
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 3. Measure the output voltage with the DMM and enter the data. This value
may be positive or negative.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 4. Select the second voltage calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should be
displayed.
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Step 6 only applies to models N6781A and N6782A.
Step 6. Repeat steps 1 through 5 for the 100 mV range.
Steps 7 through 9 only apply to model N6784A.
Step 7. Select the third voltage calibration point for the negative range.
Front Panel:
The third calibration point is automatically
selected. "Enter P3 measured data" should be
displayed.
Step 8. Measure the output voltage with the DMM and enter the data. The value
should be negative.
Front Panel:
Enter the value from the external DMM into the
data field and then select Next. When
calibration is complete, select Close.
Step 9. Repeat steps 1 through 5, 7 and 8 for the 100 mV range.
Calibration
3
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P3
*OPC?
SCPI Command:
CAL:DATA <data>
65