Calibration Procedure - Models N676Xa - Keysight N6705 Service Manual

Dc power
Table of Contents

Advertisement

Calibration Procedure – Models N676xA
Enter Calibration mode
Voltage Calibration
Model N6705 Service Guide
Warm-up Period for Models N6761A and N6762A
Models N6761A and N6762A require a 30-minute warm-up period in the
reset (*RST) state before starting calibration.
Front Panel:
Press [Menu], then select Utilities,
Administrative Tools, Administrator Login.
Enter your numeric password in the PIN field
and select Login.
Then select the Calibration menu. Select
Turn On/Off to access the menu items.
High Range Programming and Measurement
Step 1. Connect the voltage input of the Keysight 3458A to an output channel
(see Cal Setup A).
Step 2. Select the full-scale voltage programming and measurement range. The
value entered must be the maximum voltage of the range that you are
calibrating. This example selects the 50 V full-scale range of output 1.
Full-scale ranges vary according to model.
Front Panel:
From the Calibration menu, select Voltage,
High Range Prog and Meas. Then select
Start.
Step 3. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 4. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Calibration
SCPI Command:
CAL:STAT ON,
<password>
SCPI Command:
CAL:VOLT 50
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
53
3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents