Model N6705 Service Guide
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 5. Select the second current calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should be
displayed.
Step 6. Measure the current with the Keysight 3458A and enter the value. This
value should be negative.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
1 mA Measurement Range - All Models
Step 1. Connect the Keysight 3458A in series with a 10 kΩ resistor to the output
(see equipment list in chapter 2 and Cal Setup D).
Step 2. Select the 1 mA range of output 1.
Front Panel:
From the Calibration menu, select Current,
I mA Measurement Range. Select Start.
Step 3. Select the first current calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 4. Measure the current with the Keysight 3458A and enter the value.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 5. Select the second current calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should be
displayed.
Step 6. Measure the current with the Keysight 3458A and enter the value. This
value should be negative.
Calibration
3
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:CURR:MEAS
0.001
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
71