Keysight N6705 Service Manual page 78

Dc power
Table of Contents

Advertisement

3
Calibration
78
Step 6. Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
Enter the calculated value into the data field
and then select Next. When calibration is
complete, select Close.
Low Range Current Measurement
Step 1. Connect only the current measurement terminals of the Keysight 3458A
to the output terminals (see Cal Setup D without the resistor).
Step 2. Select the low current measurement range. The value to program a range
must be the maximum current of the range. This example selects the
0.15 A range of output 1.
Front Panel:
From the Calibration menu, select
Current, Low Range Measurement.
Then select Start.
Step 3. Select the first current calibration point.
Front Panel:
The first calibration point is automatically
selected. The ammeter should be connected.
Select Next. "Enter Measured Current" should
be displayed.
Step 4. Measure the current with the Keysight 3458A and enter the value.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next. When
calibration is complete, select Close.
Negative Current Limit Calibration
Step 1. Connect the Keysight 3458A, current shunt, and a power supply to the
output (see Test Setup G). Connect plus to plus, minus to minus. Set the
external power supply to 3V, 3A.
Step 2. Select the negative-current range. The value to select the range must be
negative and the maximum current of the range.
Front Panel:
From the Calibration menu, select
Current, Low Range Measurement.
Then select Start.
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:CURR:MEAS 0.15
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
*OPC?
SCPI Command:
CAL:CURR:LIM:NEG -2
Model N6705 Service Guide
, (@1)
, (@1)