3
Calibration
This chapter includes calibration procedures for the Keysight N6705 DC
Power Analyzer. Instructions are given for performing the procedures
from either the front panel or a controller over the GPIB.
Refer to the "Equipment Required" section in chapter 2 for a list of the
equipment required for calibration. Also refer to "Measurement
Techniques" for information about connecting the voltmeter and current
shunt.
Perform the verification tests before calibrating your DC Power Analyzer.
If the DC Power Analyzer passes the verification tests, the unit is
operating within its calibration limits and does not need to be re-
calibrated.
Calibration Interval
The recommended calibration interval for Keysight N6700 Series power
modules is one year. Keysight N6705 DC Power Analyzer mainframes do
not require calibration.