Appendix A
Error Messages
Device-dependent Errors (continued)
311
This group of channel has non-identical features
One or more modules in this group does not match the ratings or options of the other modules.
312
The command is disabled when the channel is grouped
This command is disabled when the output channel is grouped.
313
Datalog time interval is out of range
The data log time interval is outside the programmable range.
314
Time of day clock has stopped
The time of day clock has stopped. Replace the internal battery. See Service Guide.
315
Settings conflict error
A data element could not be programmed because of the present instrument state.
316
Mass storage error
The mass storage memory has been exceeded.
317
Invalid format
An invalid data format was found in the command string.
318
Datalog error
An error has occurred in the data returned by the datalogger.
319
Measurement is not allowed while datalogger is running
You cannot make measurements in Meter View or Scope View while the datalogger is running.
320
Firmware update error
This may be due to the instrument hardware not being able to support the firmware version.
321
Modules do not match data format
One or more power modules does not match the data format of the file that is being loaded
322
User defined arb error
An error has occurred in the format or the data of the user–defined arb.
323
Meas TINTerval too small on too many channels
The measurement time interval is too small to be supported by the number of channels.
324
Inconsistent arb settings
The arb settings are inconsistent; most likely a mismatch in the arb lengths.
325
Arb sequence error
An error has occurred with the arb sequence.
326
Not supported by emulation mode
The function/parameter is not supported by the specified emulation mode.
327
Initiated with no sense function enabled
A measurement has been initiated without specifying the measurement (sense) function.
328
Too many measurement points
Too many measurement points have been specified.
Command Errors (these errors set Standard Event Status register bit #5)
100
Command error
Generic syntax error.
101
Invalid character
An invalid character was found in the command string.
102
Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
103
Invalid separator
An invalid separator was found in the command string. Check for proper use of , ; :
104
Data type error
A different data type than the one allowed was found in the command string.
105
GET not allowed
A group execute trigger is not allowed in a command string.
128
Model N6705 Service Guide