Performance Verification - Keysight N6705 Service Manual

Dc power
Table of Contents

Advertisement

NOTE
Performance
Calibration Tests
9
2

Performance Verification

Verification Description .................................................................. 10
Verification Procedure .................................................................... 13
Test Record Forms ......................................................................... 22
The verification procedures described in this chapter verify that the
Keysight N6705 DC Power Analyzer is operating normally and is within
published specifications.
Perform the verification tests before calibrating your DC Power Analyzer.
If the DC Power Analyzer passes the verification tests, the unit is
operating within its calibration limits and does not need to be re-
calibrated.
Verification procedures perform two primary functions:
These tests verify that the DC Power Analyzer is operating normally
Tests
and meets the specifications published in the Keysight N6700 Modular
Power System Family Specifications Guide.
These tests verify that the DC Power Analyzer is operating within its
calibration limits.
If the instrument fails any of the tests or if abnormal test results are
obtained, try calibrating the unit. If calibration is unsuccessful, return the
unit to a Keysight Technologies Service Center.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents