Keysight N6705 Service Manual page 66

Dc power
Table of Contents

Advertisement

3
Calibration
66
Positive Voltage Limit - Models N6781A, N6782A, N6784A
Step 1. Select the positive voltage limit range. The value to program a range
must be the maximum voltage of the range to be calibrated. This
example selects the 20 V positive limit range of output 1.
Front Panel:
From the Calibration menu,
select Voltage, Positive Voltage
Limit. Then select Start. Select
the 20 V range.
Step 2. Select the first voltage calibration point.
Front Panel:
The first calibration point is automatically
selected. "Enter P1 measured data" should be
displayed.
Step 3. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into
the data field and then select Next.
Step 4. Select the second voltage calibration point.
Front Panel:
The second calibration point is automatically
selected. "Enter P2 measured data" should be
displayed.
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
Enter the value from the external DMM into the
data field and then select Next. When
calibration is complete, select Close.
Step 6. Repeat steps 1 through 5 for the 6 V positive limit range.
SCPI Command:
CAL:VOLT:LIM:POS 20
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
Model N6705 Service Guide
, (@1)