6.21 STORe Group
:STORe:INTerval
Function
Sets the store interval or queries the current
setting.
Syntax
:STORe:INTerval {<NRf>,<NRf>,<NRf>}
:STORe:INTerval?
1st <NRf> = 0 to 99 (hour)
2nd <NRf> = 0 to 59 (minute)
3rd <NRf> = 1 to 59 (second)
Example
:STORE:INTERVAL 0,0,0
:STORE:INTERVAL? ->
:STORE:INTERVAL 0,0,0
:STORe:ITEM
Function
Sets the stored item or queries the current
setting.
Syntax
:STORe:ITEM {NUMeric|WAVE|NWAVe}
:STORe:ITEM?
NUMeric = Store only the numeric data.
WAVE = Store only the waveform display data
NWAVe = Store both the numeric data and
waveform display data
Example
:STORE:ITEM NUMERIC
:STORE:ITEM? -> :STORE:ITEM NUMERIC
:STORe:MEMory?
Function
Queries all settings related to the storage
memory.
Syntax
:STORe:MEMory?
Example
:STORE:MEMORY? ->
:STORE:MEMORY:ALERT 1
:STORe:MEMory:ALERt
Function
Sets whether to display a confirmation message
when clearing the storage memory or queries
the current setting.
Syntax
:STORe:MEMory:ALERt {<Boolean>}
:STORe:MEMory:ALERt?
Example
:STORE:MEMORY:ALERT ON
:STORE:MEMORY:ALERT? ->
:STORE:MEMORY:ALERT 1
Description The initialization takes place immediately when
initializing the storage memory using the
":STORe:MEMory:INITialize" command
regardless of the setting specified with this
command.
:STORe:MEMory:CONVert:ABORt
Function
Abort converting the stored data from the
memory to the file.
Syntax
:STORe:MEMory:CONVert:ABORt
Example
:STORE:MEMORY:CONVERT:ABORT
6-106
:STORe:MEMory:CONVert:EXECute
Function
Executes the converting of the stored data from
the memory to the file.
Syntax
:STORe:MEMory:CONVert:EXECute
Example
:STORE:MEMORY:CONVERT:EXECUTE
Description • The convert destination file is set using the
":STORe:FILE:..." command.
• When file conversion is executed, the
WT3000 accesses the file twice.
To confirm the completion of the file
conversion, use the "COMMUNICATE:WAIT
64" command (checks the change in bit 6
(ACS) of the condition register) and check the
completion of the file access of the WT3000
twice. An example is indicated below.
"STATUS:EESR?"
(Clear the extended event register)
"STORE:MEMORY:CONVERT:EXECUTE"
(Start the file conversion)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the first
time)
"STATUS:EESR?"
(Clear the extended event register)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the second
time)
"STATUS:EESR?"
(Clear the extended event register)
:STORe:MEMory:INITialize
Function
Executes the initialization of the storage
memory.
Syntax
:STORe:MEMory:INITialize
Example
:STORE:MEMORY:INITIALIZE
:STORe:MODE
Function
Sets the data storage/recall or queries the
current setting.
Syntax
:STORe:MODE {STORe|RECall}
:STORe:MODE?
Example
:STORE:MODE STORE
:STORE:MODE? -> :STORE:MODE STORE