6.14 INPut Group
[:INPut]:CURRent:RANGe:ELEMent<x>
Function
Sets the current range of the element or queries
the current setting.
Syntax
[:INPut]:CURRent:RANGe:
ELEMent<x>
{<Current>|(EXTernal,<Voltage>)}
[:INPut]:CURRent:RANGe:ELEMent<x>?
<x> = 1 to 4 (element)
With the (standard) power element models
(760301-01 through 760304-04)
• When the crest factor is set to 3
<Current> = 500 (mA), 1, 2, 5, 10, 20, 30 (A)
(for direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5,
10 (V) (for external current sensor input)
• When the crest factor is set to 6
<Current> = 250, 500 (mA), 1, 2.5, 5, 10, 15
(A) (for direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1,
2.5, 5 (V) (for external current sensor input)
With the low current range power element
models (760301-10 through 760304-40)
• When crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500
(mA), 1, 2, (A) (with direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5,
10 (V) (with external current sensor input)
• When crest factor is set to 6
<Current> = 2.5, 5, 10, 25, 50, 100, 250, 500
(mA), 1 (A) (with direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1,
2.5, 5 (V) (with external current sensor input)
Example
:INPUT:CURRENT:RANGE:ELEMENT1 30A
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 30.0E+00
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10V
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10.00E+00
6-68
[:INPut]:CURRent:RANGe:{SIGMA|SIGMB}
Function
Collectively sets the current range of all
elements belonging to wiring unit {ΣA|ΣB}.
Syntax
[:INPut]:CURRent:RANGe:{SIGMA|
SIGMB}
{<Current>|(EXTernal,<Voltage>)}
With the (standard) power element models
(760301-01 through 760304-04)
• When the crest factor is set to 3
<Current> = 500 (mA), 1, 2, 5, 10, 20, 30 (A)
(for direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5,
10 (V) (for external current sensor input)
• When the crest factor is set to 6
<Current> = 250, 500 (mA), 1, 2.5, 5, 10, 15
(A) (for direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1,
2.5, 5 (V) (for external current sensor input)
With the low current range power element
models (760301-10 through 760304-40)
• When crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500
(mA), 1, 2, (A) (with direct current input)
<Voltage> = 50, 100, 200, 500 (mV), 1, 2, 5,
10 (V) (with external current sensor input)
• When crest factor is set to 6
<Current> = 2.5, 5, 10, 25, 50, 100, 250, 500
(mA), 1 (A) (with direct current input)
<Voltage> = 25, 50, 100, 250, 500 (mV), 1,
2.5, 5 (V) (with external current sensor input)
Example
:INPUT:CURRENT:RANGE:SIGMA 30A
Description • [:INPut]:CURRent:RANGe:SIGMA is valid
only on models with 2 to 4 elements.
• [:INPut]:CURRent:RANGe:SIGMB is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{SA|SB} does not exist as a result of the
wiring system setting ([:INPut]:WIRing).