6.16 MEASure Group
:MEASure:PC:IEC
Function
Sets the equation used to compute Pc
(Corrected Power) or queries the current
setting.
Syntax
:MEASure:PC:IEC {<NRf>}
:MEASure:PC:IEC?
<NRf> = 1976, 1993
Example
:MEASURE:PC:IEC 1976
:MEASURE:PC:IEC? ->
:MEASURE:PC:IEC 1976
Description Specify the year when the equation used to
calculate the Pc was issued by IEC76-1.
:MEASure:PC:P<x>
Function
Sets the parameter used to compute Pc
(Corrected Power) or queries the current
setting.
Syntax
:MEASure:PC:P<x> {<NRf>}
:MEASure:PC:P<x>?
<x> = 1, 2 (P1, P2)
<NRf> = 0.0001 to 9.9999
Example
:MEASURE:PC:P1 0.5
:MEASURE:PC:P1? ->
:MEASURE:PC:P1 0.5000
Description This parameter is used when the
":MEASure:PC:IEC" setting is set to
"1976(IEC76-1(1976))."
:MEASure:PHASe
Function
Sets the display format of the phase difference
or queries the current setting.
Syntax
:MEASure:PHASe {<NRf>}
:MEASure:PHASe?
<NRf> = 180 or 360
Example
:MEASURE:PHASE 180
:MEASURE:PHASE? ->
:MEASURE:PHASE 180
Description Displays the phase using ±0 to 180° (Lead/Lag)
for "180" and 0 to 360° for "360."
:MEASure:SAMPling
Function
Sets the sampling frequency or queries the
current setting.
Syntax
:MEASure:SAMPling {AUTO|CLKA|CLKB|
CLKC}
:MEASure:SAMPling?
Example
:MEASURE:SAMPLING AUTO
:MEASURE:SAMPLING? ->
:MEASURE:SAMPLING AUTO
Description For details on the sampling frequency
corresponding to {AUTO|CLKA|CLKB|CLKC},
see the User's Manual IM760301-01E .
6-80
:MEASure:SQFormula
Function
Sets the equation used to compute S (apparent
power) and Q (reactive power) or queries the
current setting.
Syntax
:MEASure:SQFormula {TYPE1|TYPE2|
TYPE3}
:MEASure:SQFormula?
Example
:MEASURE:SQFORMULA TYPE1
:MEASURE:SQFORMULA? ->
:MEASURE:SQFORMULA TYPE1
Description • For details on the equation corresponding to
{TYPE1|TYPE2|TYPE3}, see the User's
Manual IM760301-01E .
• "TYPE3" is selectable only on models with the
advanced computation function (/G6 option)
or the harmonic measurement function (/G5
option).
:MEASure:SYNChronize
Function
Sets the synchronized measurement mode or
queries the current setting.
Syntax
:MEASure:SYNChronize {MASTer|SLAVe}
:MEASure:SYNChronize?
Example
:MEASURE:SYNCHRONIZE MASTER
:MEASURE:SYNCHRONIZE? ->
:MEASURE:SYNCHRONIZE MASTER
IM 760301-17E