:FLICker:PLT:LIMit
Function
Sets the limit of the long-term flicker value Plt or
queries the current setting.
Syntax
:FLICker:PLT:LIMit {<NRf>}
:FLICker:PLT:LIMit?
<NRf> = 0.10 to 99.99 (limit)
Example
:FLICKER:PLT:LIMIT 0.65
:FLICKER:PLT:LIMIT? ->
:FLICKER:PLT:LIMIT 0.65
:FLICker:PLT:NVALue
Function
Sets constant N for the equation used to
compute the long-term flicker value Plt or
queries the current setting.
Syntax
:FLICker:PLT:NVALue {<NRf>}
:FLICker:PLT:NVALue?
<NRf> = 1 to 99 (constant N)
Example
:FLICKER:PLT:NVALUE 12
:FLICKER:PLT:NVALUE? ->
:FLICKER:PLT:NVALUE 12
:FLICker:PLT[:STATe]
Function
Turns ON/OFF judgment of the long-term flicker
value Plt or queries the current setting.
Syntax
:FLICker:PLT[:STATe] {<Boolean>}
:FLICker:PLT:STATe?
Example
:FLICKER:PLT:STATE ON
:FLICKER:PLT:STATE? ->
:FLICKER:PLT:STATE 1
:FLICker:PST?
Function
Queries all settings related to the short-term
flicker value Pst.
Syntax
:FLICker:PST?
Example
:FLICKER:PST? ->
:FLICKER:PST:STATE 1;LIMIT 1.00
:FLICker:PST:LIMit
Function
Sets the limit for the short-term flicker value or
queries the current setting.
Syntax
:FLICker:PST:LIMit {<NRf>}
:FLICker:PST:LIMit?
<NRf> = 0.10 to 99.99 (limit)
Example
:FLICKER:PST:LIMIT 1.00
:FLICKER:PST:LIMIT? ->
:FLICKER:PST:LIMIT 1.00
IM 760301-17E
:FLICker:PST[:STATe]
Function
Turns ON/OFF judgment of the short-term
flicker value Pst or queries the current setting.
Syntax
:FLICker:PST[:STATe] {<Boolean>}
:FLICker:PST:STATe?
Example
:FLICKER:PST:STATE ON
:FLICKER:PST:STATE? ->
:FLICKER:PST:STATE 1
:FLICker:RESet
Function
Resets measured flicker data.
Syntax
:FLICker:RESet
Example
:FLICKER:RESET
:FLICker:STARt
Function
Starts flicker measurement.
Syntax
:FLICker:STARt
Example
:FLICKER:START
:FLICker:STATe?
Function
Queries the status of flicker measurement.
Syntax
:FLICker:STATe?
Example
:FLICKER:STATE? -> RESET
Description The contents of the response are as follows:
RESet = Reset status
INITialize = Initializing
READy = Measurement start wait state
STARt = Measuring
COMPlete = Measurement stopped, judgment
results displayed
:FLICker:UN?
Function
Queries all settings related to rated voltage Un.
Syntax
:FLICker:UN?
Example
:FLICKER:UN? -> :FLICKER:UN:
MODE AUTO;VALUE 230.00
:FLICker:UN:MODE
Function
Sets the assignment method for rated voltage
Un or queries the current setting.
Syntax
:FLICker:UN:MODE {AUTO|SET}
:FLICker:UN:MODE?
AUTO = Use the voltage value measured upon
start of measurement.
SET = Use the predefined value
(:FLICker:UN:VALue).
Example
:FLICKER:UN:MODE AUTO
:FLICKER:UN:MODE? ->
:FLICKER:UN:MODE AUTO