6.8 FILE Group
:FILE:SAVE:NUMeric:NORMal?
Function
Queries all settings related to the items saved to
numeric data files.
Syntax
:FILE:SAVE:NUMeric:NORMal?
Example
:FILE:SAVE:NUMERIC:NORMAL? ->
:FILE:SAVE:NUMERIC:NORMAL:
ELEMENT1 1;ELEMENT2 0;ELEMENT3 0;
ELEMENT4 0;SIGMA 0;SIGMB 0;U 1;I 1;
P 1;S 1;Q 1;LAMBDA 1;PHI 1;FU 1;
FI 1;UPPEAK 0;UMPEAK 0;IPPEAK 0;
IMPEAK 0;CFU 0;CFI 0;PC 0;TIME 0;
WH 0;WHP 0;WHM 0;AH 0;AHP 0;AHM 0;
WS 0;WQ 0;ETA1 0;ETA2 0;ETA3 0;
ETA4 0;F1 0;F2 0;F3 0;F4 0;F5 0;
F6 0;F7 0;F8 0;F9 0;F10 0;F11 0;
F12 0;F13 0;F14 0;F15 0;F16 0;
F17 0;F18 0;F19 0;F20 0
:FILE:SAVE:NUMeric[:NORMal]:ALL
Function
Collectively turns ON/OFF the output of all
element functions when saving the numerical
data file.
Syntax
:FILE:SAVE:NUMeric[:NORMal]:
ALL {<Boolean>}
Example
:FILE:SAVE:NUMERIC:NORMAL:ALL ON
:FILE:SAVE:NUMeric[:NORMal]:{ELEMent<x>|
SIGMA|SIGMB}
Turns ON/OFF the output of {each
Function
element|ΣA|ΣB} when saving the numeric data
to file.
Syntax
:FILE:SAVE:NUMeric[:NORMal]:
{ELEMent<x>|SIGMA|
SIGMB} {<Boolean>}
:FILE:SAVE:NUMeric[:NORMal]:
{ELEMent<x>|SIGMA|SIGMB}?
<x> = 1 to 4
Example
:FILE:SAVE:NUMERIC:NORMAL:
ELEMENT1 ON
:FILE:SAVE:NUMERIC:NORMAL:ELEMENT1?
-> :FILE:SAVE:NUMERIC:NORMAL:
ELEMENT1 1
Description • :FILE:SAVE:NUMeric[:NORMal]:SIGMA
is valid on models with two or more elements.
To turn the output ON, wiring unit ΣA must
exist by setting the wiring system beforehand
using the [:INPut]WIRing command.
• :FILE:SAVE:NUMeric[:NORMal]:SIGMB
is valid on models with four elements. To turn
the output ON, wiring unit ΣB must exist by
setting the wiring system beforehand using
the [:INPut]WIRing command.
6-50
:FILE:SAVE:NUMeric[:NORMal]:PRESet<x>
Function
Presets the output ON/OFF pattern of the
element function for saving the numeric data to
file.
Syntax
:FILE:SAVE:NUMeric[:NORMal]:
PRESet<x>
<x> = 1 to 2 (preset pattern number)
Example
:FILE:SAVE:NUMERIC:NORMAL:PRESET1
Description For details on the output pattern when preset is
executed, see the User's Manual IM760301-
01E .
:FILE:SAVE:NUMeric[:NORMal]:<Function>
Function
Turns ON/OFF the output of the function when
saving the numerical data file or queries the
current setting.
Syntax
:FILE:SAVE:NUMeric[:NORMal]:
<Function> {<Boolean>}
:FILE:SAVE:NUMeric[:NORMal]:
<Function>?
<Function> = {U|I|P|S|Q|...}(See the
function selection list (1) of "DISPlay group" on
page 6-44.)
Example
:FILE:SAVE:NUMERIC:NORMAL:U ON
:FILE:SAVE:NUMERIC:NORMAL:U? ->
:FILE:SAVE:NUMERIC:NORMAL:U 1
:FILE:SAVE:NUMeric:TYPE
Function
Sets the format of the numeric data to be saved
or queries the current setting.
Syntax
:FILE:SAVE:NUMeric:TYPE {ASCii|
FLOat}
:FILE:SAVE:NUMeric:TYPE?
Example
:FILE:SAVE:NUMERIC:TYPE ASCII
:FILE:SAVE:NUMERIC:TYPE? ->
:FILE:SAVE:NUMERIC:TYPE ASCII
:FILE:SAVE:SETup[:EXECute]
Function
Saves of the setup parameter file.
Syntax
:FILE:SAVE:SETup[:EXECute]
{<Filename>}
Example
:FILE:SAVE:SETUP:EXECUTE "SETUP1"
Description • Specify the file name without the extension.
• This command is an overlap command.
:FILE:SAVE:WAVE?
Function
Queries all settings related to the saving of
waveform display data files.
Syntax
:FILE:SAVE:WAVE?
Example
:FILE:SAVE:WAVE? ->
:FILE:SAVE:WAVE:TYPE BINARY
IM 760301-17E

Advertisement

Table of Contents
loading

Table of Contents