YOKOGAWA WT3000 User Manual page 151

Precision power analyzer communication interface
Hide thumbs Also See for WT3000:
Table of Contents

Advertisement

6.21 STORe Group
:STORe:RECall
Function
Sets the data number to be recalled or queries
the current setting.
Syntax
:STORe:RECall {<NRf>}
:STORe:RECall?
<NRf> = 1 to 999999
Example
:STORE:RECALL 1
:STORE:RECALL? -> :STORE:RECALL 1
:STORe:RTIMe?
Function
Queries the store reservation time for real-time
store mode.
Syntax
:STORe:RTIMe?
Example
:STORE:RTIME? -> :STORE:RTIME:
START 2005,1,1,0,0,0;
END 2005,1,1,1,0,0
:STORe:RTIMe:{STARt|END}
Function
Sets the store {start|stop} reservation date/time
for real-time store mode or queries the current
setting.
Syntax
:STORe:RTIMe:{STARt|END} {<NRf>,
<NRf>,<NRf>,<NRf>,<NRf>,<NRf>}
:STORe:RTIMe:{STARt|END}?
{<NRf>, <NRf>, <NRf>, <NRf>, <NRf>, <NRf>}
= 2001, 1, 1, 0, 0, 0 to 2099, 12, 31, 23, 59, 59
1st <NRf> = 2001 to 2099 (year)
2nd <NRf> = 1 to 12 (month)
3rd <NRf> = 1 to 31 (day)
4th <NRf> = 0 to 23 (hour)
5th <NRf> = 0 to 59 (minute)
6th <NRf> = 0 to 59 (second)
Example
:STORE:RTIME:START 2005,1,1,0,0,0
:STORE:RTIME:START? ->
:STORE:RTIME:START 2005,1,1,0,0,0
Description This command is valid when the store mode
(:STORe:SMODe) is set to RTIMe (real-time
store mode).
:STORe:SMODe
Function
Sets the store mode or queries the current
setting.
Syntax
:STORe:SMODe {MANual|RTIMe|
INTEGrate}
:STORe:SMODe?
MANual = Manual store mode
RTIMe = Real-time store mode
INTEGrate = Integration synchronization store
mode
Example
:STORE:SMODE MANUAL
:STORE:SMODE? ->
:STORE:SMODE MANUAL
6-108
:STORe:STARt
Function
Starts the data store operation.
Syntax
:STORe:STARt
Example
:STORE:START
Description When ":STORe:SMODe" is set to MANual, the
storage operation is executed. When set to
{RTIMe|INTEGrate} the WT3000 enters the
store wait state.
:STORe:STOP
Function
Stops the data storage operation.
Syntax
:STORe:STOP
Example
:STORE:STOP
:STORe:WAVE?
Function
Queries all settings related to the storage of
waveform display data.
Syntax
:STORe:WAVE?
Example
:STORE:WAVE? -> :STORE:WAVE:U1 1;
U2 0;U3 0;U4 0;I1 1;I2 0;I3 0;I4 0
:STORe:WAVE:ALL
Function
Collectively turns ON/OFF the output of all
waveforms when storing waveform display data.
Syntax
:STORe:WAVE:ALL {<Boolean>}
Example
:STORE:WAVE:ALL ON
:STORe:WAVE:{U<x>|I<x>|SPEed|TORQue}
Function
Turns ON/OFF the output of the waveform when
storing the waveform display data or queries the
current setting.
Syntax
:STORe:WAVE:{U<x>|I<x>|SPEed|
TORQue} {<Boolean>}
:STORe:WAVE:{U<x>|I<x>|SPEed|
TORQue}?
<x> = 1 to 4
Example
:STORE:WAVE:U1 ON
:STORE:WAVE:U1? -> :STORE:WAVE:U1 1
Description {SPEed|TORQue} are valid only on the motor
version (-MV).
IM 760301-17E

Advertisement

Table of Contents
loading

Table of Contents