YOKOGAWA WT3000 User Manual page 68

Precision power analyzer communication interface
Hide thumbs Also See for WT3000:
Table of Contents

Advertisement

:COMMunicate:STATus?
Function
Queries line-specific status.
Syntax
:COMMunicate:STATus?
Example
:COMMUNICATE:STATUS? ->
:COMMUNICATE:STATUS 0
Description The meaning of each status bit is as follows:
Bit
GP-IB
0
Unrecoverable
transmission error
1
Always 0
2
Always 0
3 or greater Always 0
The value 0 is always returned for the optional
USB or Ethernet interface.
The status bit is set when the corresponding
cause occurs and cleared when it is read.
:COMMunicate:VERBose
Function
Sets whether to return the response to a query
using full spelling (example
:INPUT:VOLTAGE:RANGE:ELEMENT1
1.000E+03) or using abbreviation (example
:VOLT:RANG:ELEM 1.000E+03).
Syntax
:COMMunicate:VERBose {<Boolean>}
:COMMunicate:VERBose?
Example
:COMMUNICATE:VERBOSE ON
:COMMUNICATE:VERBOSE? ->
:COMMUNICATE:VERBOSE 1
:COMMunicate:WAIT
Function
Waits for one of the specified extended events
to occur.
Syntax
:COMMunicate:WAIT <Register>
<Register> = 0 to 65535 (extended event
register, see page 7-7.)
Example
:COMMUNICATE:WAIT 1
Description For the description regarding how to
synchronize the program using
COMMunicate:WAIT, see page 5-9.
:COMMunicate:WAIT?
Function
Creates the response that is returned when the
specified event occurs.
Syntax
:COMMunicate:WAIT? <Register>
<Register> = 0 to 65535 (extended event
register, see page 7-7.)
Example
:COMMUNICATE:WAIT? 65535 -> 1
Operation pending status register/Overlap
enable register
15
14
13
12
11
10
0
0
0
0
0
0
When bit 5 (PRN) = 1:
Built-in printer operation not complete
When bit 6 (ACS) = 1:
Access to the medium not complete.
IM 760301-17E
RS-232
Parity error
Framing error
Break character
detected
Always 0
9
8
7
6
5
4
3
2
1
0
0
0
0 ACS
PRN
0
0
0
0
0
6.5 COMMunicate Group
6-25
1
2
3
4
5
6
7
App
Index

Advertisement

Table of Contents
loading

Table of Contents