:DISPlay:MODE
Function
Sets the display mode or queries the current
setting.
Syntax
:DISPlay:MODE {NUMeric|WAVE|BAR|
VECTor|TRENd|NWAVe|NBAR|NTRend|
WBAR|WTRend|BTRend|MATH|NMATh|FFT|
NFFT|WFFT|IECHarm|CBCycle|FLICker}
:DISPlay:MODE?
NUMeric = Displays only the numeric values.
WAVE = Displays only the waveforms.
BAR = Bar graph
VECTor = Vector display
TRENd = Trend
NWAVe = Displays both the numeric values and
the waveforms.
NBAR = Displays both the numeric values and
the bar graph.
NTRend = Displays both the numeric values
and the trends.
WBAR = Displays both the waveforms and the
bar graph.
WTRend = Displays both the waveforms and
the trends.
BTRend = Displays both the bar graphs and the
trends.
MATH = Displays only the waveforms (including
the computed waveforms).
NMATh = Displays both numeric values and
waveforms (including the computed
waveforms).
FFT = Displays only the FFT waveforms.
NFFT = Displays both the numeric values and
the FFT waveforms.
WFFT = Displays both the waveforms and the
FFT waveforms.
IECHarm = IEC harmonic measurement mode
display (numeric values).
CBCycle = Cycle by Cycle mode display (value).
FLICker = Flicker measurement mode display
(value).
Example
:DISPLAY:MODE NUMERIC
:DISPLAY:MODE? ->
:DISPLAY:MODE NUMERIC
Description • {BAR|VECTor|NBAR|WBAR|BTRend} are
selectable only on models with the advanced
computation function (/G6 option) or the
harmonic measurement function (/G5 option).
• {MATH|NMATh|FFT|NFFT|WFFT|
IECHarm} are selectable only on models with
the advanced computation function (/G6
option).
• {CBCycle} can only be selected with the
Cycle by Cycle function (/CC option).
• {FLICker} can only be selected with the
flicker measurement function (/FL option).
IM 760301-17E
:DISPlay:NUMeric?
Function
Queries all settings related to the numeric
display.
Syntax
:DISPlay:NUMeric?
Example
:DISPLAY:NUMERIC? -> (same as the
response to ":DISPlay:NUMeric:NORMal?")
:DISPlay:NUMeric:NORMal?
Function
Queries all settings related to the numeric
display.
Syntax
:DISPlay:NUMeric:NORMal?
Example
• Example in which the numeric display format
(:DISPlay:NUMeric[:NORMal]:FORMat)
is set to "VAL4 (4-value display)"
:DISPLAY:NUMERIC:NORMAL? ->
:DISPLAY:NUMERIC:NORMAL:
FORMAT VAL4;VAL4:ITEM1 U,1,TOTAL;
ITEM2 I,1,TOTAL;ITEM3 P,1,TOTAL;
...(omitted)...;ITEM35 ETA3;
ITEM36 ETA4;CURSOR 1
• Example in which the numeric display format
(:DISPlay:NUMeric[:NORMal]:FORMat)
is set to "ALL (all display)"
:DISPLAY:NUMERIC:NORMAL? ->
:DISPLAY:NUMERIC:NORMAL:
FORMAT ALL;ALL:CURSOR U
Description Returns all settings corresponding to the
numeric display format
(:DISPlay:NUMeric[:NORMal]:FORMat).
:DISPlay:NUMeric[:NORMal]:ALL?
Function
Queries all settings related to the numeric
display (all display).
Syntax
:DISPlay:NUMeric[:NORMal]:ALL?
Example
:DISPLAY:NUMERIC:NORMAL:ALL? ->
:DISPLAY:NUMERIC:NORMAL:ALL:
CURSOR U
:DISPlay:NUMeric[:NORMal]:ALL:CURSor
Function
Sets the cursor position on the numeric display
(all display) or queries the current setting.
Syntax
:DISPlay:NUMeric[:NORMal]:ALL:
CURSor {<Function>}
:DISPlay:NUMeric[:NORMal]:ALL:
CURSor?
<Function> = {U|I|P|S|Q|...} (See the
function selection list (1) on page 6-44.)
Example
:DISPLAY:NUMERIC:NORMAL:ALL:
CURSOR U
:DISPLAY:NUMERIC:NORMAL:ALL:CURSOR?
-> :DISPLAY:NUMERIC:NORMAL:ALL:
CURSOR U
Description Specify the cursor position in terms of the
function name.