YOKOGAWA WT3000 User Manual page 74

Precision power analyzer communication interface
Hide thumbs Also See for WT3000:
Table of Contents

Advertisement

:DISPlay:FFT:FFT<x>?
Function
Queries all settings related to the FFT
waveform.
Syntax
:DISPlay:FFT:FFT<x>?
<x> = 1, 2 (FFT)
Example
:DISPLAY:FFT:FFT1? -> :DISPLAY:FFT:
FFT1:STATE 1;OBJECT U1;LABEL "FFT1"
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:FFT:FFT<x>:LABel
Function
Sets the label of the FFT waveform or queries
the current setting.
Syntax
:DISPlay:FFT:FFT<x>:
LABel {<String>}
:DISPlay:FFT:FFT<x>:LABel?
<x> = 1, 2 (FFT)
<String> = Up to 8 characters
Example
:DISPLAY:FFT:FFT1:LABEL "FFT1"
:DISPLAY:FFT:FFT1:LABEL? ->
:DISPLAY:FFT:FFT1:LABEL "FFT1"
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:FFT:FFT<x>:OBJect
Function
Sets the source waveform of the FFT
computation or queries the current setting.
Syntax
:DISPlay:FFT:FFT<x>:OBJect {U<x>|
I<x>|P<x>|PA|PB|Q<x>|QA|QB|SPEed|
TORQue}
:DISPlay:FFT:FFT<x>:OBJect?
<x> of FFT<x> = 1, 2 (FFT)
<x> of U<x>, I<x>, P<x>, Q<x> = 1 to 4
(element)
PA, QA = PΣA, QΣA (only on models with 2 to 4
elements)
PB, QB = PΣB, QΣB (only on models with 4
elements)
Example
:DISPLAY:FFT:FFT1:OBJECT U1
:DISPLAY:FFT:FFT1:OBJECT? ->
:DISPLAY:FFT:FFT1:OBJECT U1
Description • This command is valid only on models with
the advanced computation function (/G6
option).
• {SPEed|TORQue} are valid only on the
motor version (-MV).
IM 760301-17E
:DISPlay:FFT:FFT<x>[:STATe]
Function
Turns ON/OFF the FFT waveform display or
queries the current setting.
Syntax
:DISPlay:FFT:
FFT<x>[:STATe] {<Boolean>}
:DISPlay:FFT:FFT<x>:STATe?
<x> = 1, 2 (FFT)
Example
:DISPLAY:FFT:FFT1:STATE ON
:DISPLAY:FFT:FFT1:STATE? ->
:DISPLAY:FFT:FFT1:STATE 1
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:FFT:FORMat
Function
Sets the display format of the FFT waveform or
queries the current setting.
Syntax
:DISPlay:FFT:FORMat {SINGle|DUAL}
:DISPlay:FFT:FORMat?
Example
:DISPLAY:FFT:FORMAT SINGLE
:DISPLAY:FFT:FORMAT? ->
:DISPLAY:FFT:FORMAT SINGLE
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:FFT:POINt
Function
Sets the number of points of the FFT
computation or queries the current setting.
Syntax
:DISPlay:FFT:POINt {<NRf>}
:DISPlay:FFT:POINt?
<NRf> = 20000, 200000
Example
:DISPLAY:FFT:POINT 20000
:DISPLAY:FFT:POINT? ->
:DISPLAY:FFT:POINT 20000
Description This command is valid only on models with the
advanced computation function (/G6 option).
:DISPlay:FFT:SCOPe
Function
Sets the display range of the FFT waveform or
queries the current setting.
Syntax
:DISPlay:FFT:SCOPe {<NRf>,<NRf>}
:DISPlay:FFT:SCOPe?
1st <NRf> = 0 to 99990 (display start point)
2nd <NRf> = 10 to 100000 (display end point)
Example
:DISPLAY:FFT:SCOPE 0,10000
:DISPLAY:FFT:SCOPE? ->
:DISPLAY:FFT:SCOPE 0,10000
Description • This command is valid only on models with
the advanced computation function (/G6
option).
• Set the start point and then the end point.
• Set the end point so that it is greater than or
equal to (start point + 10).
6.7 DISPlay Group