6.16 MEASure Group

The commands in this group deal with computation.
You can make the same settings and inquiries as when MEASURE, AVG, "Frequency Meas. Item" menu of ITEM,
and "η Formula," "Compensation," and "∆ Measure" menus of WIRING on the front panel are used.
:MEASure?
Function
Queries all settings related to the computation.
Syntax
:MEASure?
Example
:MEASURE? -> :MEASURE:AVERAGING:
STATE 0;TYPE EXPONENT;COUNT 2;:
MEASURE:FREQUENCY:ITEM1 U1;
ITEM2 I1;:MEASURE:SAMPLING AUTO;
SQFORMULA TYPE1;PC:IEC 1976;
P1 0.5000;P2 0.5000;:MEASURE:
EFFICIENCY:ETA1 PB,PA;ETA2 PA,PB;
ETA3 OFF;ETA4 OFF;UDEF1 P1;
UDEF2 P1;:MEASURE:FUNCTION1:
STATE 0;EXPRESSION "UMN(E1)";
UNIT "V";:MEASURE:FUNCTION2:
STATE 0;EXPRESSION "UMN(E2)";
UNIT "V";:MEASURE:FUNCTION3:
STATE 0;EXPRESSION "UMN(E3)";
UNIT "V";:MEASURE:FUNCTION4:
STATE 0;EXPRESSION "UMN(E4)";
UNIT "V";:MEASURE:FUNCTION5:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION6:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION7:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION8:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION9:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION10:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION11:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION12:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION13:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION14:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION15:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION16:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION17:
STATE 0;EXPRESSION "U(E1,ORT)";
UNIT "V";:MEASURE:FUNCTION18:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:FUNCTION19:
STATE 0;EXPRESSION "U(E1,ORT)";
6-76
UNIT "V";:MEASURE:FUNCTION20:
STATE 0;EXPRESSION "I(E1,ORT)";
UNIT "A";:MEASURE:PHASE 180;
SYNCHRONIZE MASTER;MHOLD 0;
COMPENSATION:WIRING:ELEMENT1 OFF;
ELEMENT2 OFF;ELEMENT3 OFF;
ELEMENT4 OFF;:MEASURE:COMPENSATION:
EFFICIENCY 0
:MEASure:AVERaging?
Function
Queries all settings related to averaging.
Syntax
:MEASure:AVERaging?
Example
:MEASURE:AVERAGING? ->
:MEASURE:AVERAGING:STATE 0;
TYPE EXPONENT;COUNT 2
:MEASure:AVERaging:COUNt
Function
Sets the averaging coefficient or queries the
current setting.
Syntax
:MEASure:AVERaging:COUNt {<NRf>}
:MEASure:AVERaging:COUNt?
<NRf> = 2, 4, 8, 16, 32, 64 (attenuation constant
when TYPE = EXPonent)
<NRf> = 8, 16, 32, 64, 128, or 256
average count when TYPE = LINear)
Example
:MEASURE:AVERAGING:COUNT 2
:MEASURE:AVERAGING:COUNT? ->
:MEASURE:AVERAGING:COUNT 2
Description The averaging of measurement functions of
harmonic measurement (option) is valid only
when TYPE = EXPonent (attenuation constant).
For details, see the User's Manual IM760301-
01E .
:MEASure:AVERaging[:STATe]
Function
Turns ON/OFF averaging or queries the current
setting.
Syntax
:MEASure:AVERaging
[:STATe] {<Boolean>}
:MEASure:AVERaging:STATe?
Example
:MEASURE:AVERAGING:STATE ON
:MEASURE:AVERAGING:STATE? ->
:MEASURE:AVERAGING:STATE 1