Chapter 5B - Corrective Maintenance - Pca Module; Introduction - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

WARNING
Disconnect the instrument from the Alaris
System before performing maintenance. Only
qualified personnel using proper grounding
techniques should open the instrument case.
CAUTION
devices are sensitive to static electrical
CMOS
charges and may be damaged during repair if the
repair activity is not performed in an electrostatic
discharge (
) protected environment using
ESD
approved
protective procedures, including
ESD
personnel grounding.
NOTE: Due to product changes over time,
components/assemblies illustrated in this chapter
may differ from the disassembled instrument. If
there are any questions, contact CareFusion
Technical Support.
Chapter 5B — CORRECTIVE MAINTENANCE
5.1
®

INTRODUCTION

This chapter describes how to disassemble
and reassemble the PCA Module.
For more efficient repair, read the "Principles
of Operation" chapter for information on the
mechanical and electrical functions of the
PCA Module.
The surface mount devices on the circuit
boards are not field repairable. Return
circuit boards to an authorized CareFusion
Service Center for repair. Attempting
circuit board repairs voids all warranties.
Whenever the PCA Module is repaired,
perform the minimum tests required for new
instrument checkout using Maintenance
Software.
For information on replacement parts, see
the "Illustrated Parts Breakdown" chapter.
Following any level of maintenance, perform
the applicable tests (refer to "Level of
Testing Guidelines" table).
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
PCA Module
5B-1