CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.5 Removing Rear Case

1. Remove four (
washers) attaching Rear Case to
chassis assembly.
2. Pull Rear Case away from chassis
assembly.
During Reassembly:
• Ensure that Silicone Tubing gasket in
Rear Case is in place and undamaged
(see Figure
7B-10
5B-8
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screws (and associated
4
).
Lock Assembly Harness
• Raise Drive Head to highest position
before sliding Chassis Assembly into
Rear Case.
• Slide edges of Rear Case into Wiper
Assembly flange.
• Route Lock Assembly harness along
flange on Rear Case.
Rear Case Assembly
Drive Head
Slide edge of Rear Case
into Wiper Assembly
flange at reassembly