CareFusion Alaris 8110 series Service Manual page 110

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

TROUBLESHOOTING
Technical Troubleshooting Guide (Continued)
Table 6-1.
Message
Pressure Disc Installed
(Syringe Module only)
Pressure Disc Removed
(Syringe Module only)
Syringe Calibration
6-4
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
Probable Cause
1. Pressure sensing disc installed
during an infusion.
2. Sensor not sensing disc.
1. Pressure sensing disc removed
during an infusion.
2. Sensor is sensing disc.
1. System detected Force Sensor
Board, pressure disc, or Barrel
Clamp requires calibration.
2. Force Sensor Board, pressure disc,
Syringe Size Sensor Assembly, or
Logic Board fails calibration.
Remedy
1a. Press
key.
RESTART
1b. Check for proper pressure sensing
disc installation. Disc must snap into
holder.
2. If pressure sensing disc is not
installed, check binary sensor by
inserting and removing disc.
Use Maintenance Software to
perform Binary Switches Test.
Replace Pressure Sensor Board,
Pressure Sensor Harness, or Flag
Disk Holder if necessary.
3. Return to factory.
1a. Reinsert pressure sensing disc and
press
key.
RESTART
1b. Check for proper pressure sensing
disc installation. Disc must snap into
holder.
2. If pressure sensing disc is installed,
check binary sensor by inserting and
removing disc.
Use Maintenance Software to
perform Binary Switches Test.
Replace Pressure Sensor Board,
Pressure Sensor Harness, or Flag
Disk Holder if necessary.
3. Return to factory.
1. Use Maintenance Software to
perform applicable calibration(s) or a
full calibration.
2. Replace part(s) failing calibration.
Perform applicable calibration and
verification.
3. Return to factory.