DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.5 Removing Flex Cable Retainer

1. Peel Flex Cable Retainer from Internal
Frame.
2. Use a cotton swab or soft cloth lightly
dampened with
remove adhesive debris from Internal
Frame.
Retainer tucks into
Front Case here.
Isopropyl Alcohol to
70%
Retainer adheres to
Internal Frame here.
CORRECTIVE MAINTENANCE
During Reassembly:
1. Ensure Internal Frame surface (where
retainer is adhered) is clean and free of
adhesive debris.
2. Remove protective backing from
replacement retainer and adhere
retainer to Internal Frame.
3. If necessary, bend retainer on each fold
line at a
angle, toward front of
90°
instrument.
4. Lay retainer over Flex Cable and tuck
inside Front Case.