CORRECTIVE MAINTENANCE
Syringe Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2
5.2.14 Removing Syringe Size Sensor
Assembly
1. Remove one (
Size Sensor Assembly and Shaft
Support to Front Case.
2. Remove Syringe Size Sensor Assembly.
Potentiometer Extension Guide
Syringe Size Sensor Assembly
5A-16
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
) screw attaching Syringe
1
During Reassembly:
1. Insert tab on Syringe Size Sensor
Assembly into Potentiometer Extension
Guide.
2. Secure Syringe Size Sensor Assembly
to Front Case.