TROUBLESHOOTING
Error Codes (Continued)
Table 6-4.
Error Code
352.6700
353.6650
354.6670
354.6690
355.6620
356.6710
356.6780
Replacing the Force Sensor Assembly or Logic Board Assembly in a module with version
must be performed at the depot. Refer to the note at the beginning of this chapter for further information.
6-8
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
Subsystem
Logic Board
Force Sensor
Linear Position Sensor
Pressure Sensing Disc
(Syringe Module only)
Logic Board
Syringe Sizer Sensor
Digital Sensor
PCA Handset
Explanation
Force sensor circuit check
failure.
Linear position sensor output
is zero or is outside
acceptable range.
Pressure sensor output is
outside acceptable range.
Pressure Sensor current
monitor: current is outside
acceptable range.
Syringe sizer output is outside
acceptable range.
Digital sensor is outside
acceptable range during no-
power test.
System detected PCA
Handset Key stuck at power-
up.
Response
If error is repeatable: Replace
Logic Board or Force Sensor
Assembly.
If error is repeatable: Replace
linear position sensor.
Replace pressure sensing
disc.
If error is repeatable: Replace
Logic Board or Pressure
Sensor.
If error is repeatable: Replace
syringe sizer sensor.
If error is repeatable: Replace
digital sensor.
If error is repeatable: Replace
PCA Handset.
or earlier software
7