CORRECTIVE MAINTENANCE
Syringe Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2
5.2.12 Removing Top Disk Holder
CAUTION
To replace "Older Version" Top Disk Holder
(see below), replace with Front Case
Assembly Kit (Table
1. Remove two (
Disk Holder to Front Case.
2. Remove Top Disk Holder.
5A-14
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
, item
).
7A-1
9
) screws attaching Top
2
NOTE: When replacing the Top Disk Holder,
remove the Flag Disk Holder and Spring from
the old part and reuse.
During Reassembly:
• Set spring torque by inserting end of
Spring in hole of Flag Disk Holder and
winding holder counterclockwise.
• Ensure Silicone Tubing and Gasket in
Top Disk Holder mounting area on Front
Case are in place and undamaged. (See
Figure
7A
Top Disk Holder
Spring
Flag Disk Holder
-
)
5
"Older Version"
Top Disk Holder
See
CAUTION
statement above.