2.4
CONFIGURATION SETUP - SYRINGE
MODULE (Continued)
2.4.6 Max rate (Continued)
System Configuration - Syringe
Maximum rate
adjustment
Maximum
rate:
99.9
3. To accept change, press Confirm soft
key.
• Setting changes and System Config
- Syringe page

2.4.7 Near End

Enabling Near-End-of-Infusion (
allows an alert to be set from
minutes before the infusion is complete.
60
The alert occurs at the configured time or
when
of the
25%
occurs later.
NOTE: In the following procedure, the Near
End option is changed to Enable. Changing
the option setting to Disable disables the
Change Time soft key, as displayed in
step 1 .
1. After accessing System Config -
Syringe options display, page
Near End soft key.
Change
Value
mL/h
Confirm
displays.
2
NEOI
to
1
remains, whichever
VTBI
, press
2
2. To change option setting, press soft key
next to applicable option (Enable or
Disable).
)
3. To change Alert Time, press Change
Time soft key and enter new time
(
1
4. To accept changes(s), press
soft key.
• Settings change and System Config
CHECKOUT AND CONFIGURATION
System Configuration - Syringe
Enable
Near-End-of
Infusion
Disable
Alert Time:
Change
2 min
Time
Allows use of the Near-End-of
Infusion feature.
Disable
Disable
CONFIRM
System Configuration - Syringe
Enable
Near-End-of
Infusion
Disable
Alert Time:
Change
2 min
Time
Allows use of the Near-End-of
Infusion feature.
Disable
Disable
CONFIRM
to
minutes).
60
System Configuration - Syringe
Enable
Near-End-of
Infusion
Disable
Alert Time:
Change
10 min
Time
Allows use of the Near-End-of
Infusion feature.
Disable
Disable
CONFIRM
- Syringe page
displays.
2
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
CONFIRM
2-7