CareFusion Alaris 8110 series Service Manual page 89

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

DISASSEMBLY/REASSEMBLY (Continued)
5.2
5.2.14 Removing Syringe Size Sensor
Assembly
1. Remove one (
Size Sensor Assembly and Shaft
Support to Front Case.
2. Remove Syringe Size Sensor Assembly.
Syringe Size
Sensor Assembly
) screw attaching Syringe
1
Potentiometer
Extension Guide
CORRECTIVE MAINTENANCE
During Reassembly:
1. Insert tab on Syringe Size Sensor
Assembly into Potentiometer Extension
Guide.
2. Secure Syringe Size Sensor Assembly
to Front Case.
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
PCA Module
Front Case
5B-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Alaris 8120 series

Table of Contents

Save PDF