Level of Testing Guidelines - PCA Module (Continued)
Table 5B-3.
Tests to Perform
= Required
X = If verification fails.
Blank = Not Applicable
Repair/Replacement of:
Syringe Barrel Clamp assy.
Syringe Size Sensor Assy.
Miscellaneous:
Instrument dropped
Instrument software flashed
New instrument checkout
No fault found
(instrument not opened)
No fault found
(instrument opened)
NOTES:
• Use the Maintenance Software to perform testing and calibration.
• After performing any calibration procedure, always perform the
associated verification test.
CORRECTIVE MAINTENANCE
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
PCA Module
5B-31