CHECKOUT AND CONFIGURATION
2.4
CONFIGURATION SETUP - SYRINGE
MODULE (Continued)
2.4.5 KVO (Continued)
3. To change
KVO
Rate soft key and enter new rate (
mL/h in
2.5
0.01
NOTE: Flow rates less than 0.01 mL/h
are available only with a 1 cc syringe.
For larger syringes, the lower limit
adjusts to 0.1 mL/h.
System Configuration - Syringe
KVO
Rate:
_ 2
mL/h
Volume:
2%
Allows use of the KVO feature.
4. To change
KVO
Volume soft key and enter new volume
level (
to
0.5
System Configuration - Syringe
KVO
Rate:
2
mL/h
Volume:
_5%
Allows use of the KVO feature.
5. To accept changes(s), press
soft key.
• Settings change and System Config
- Syringe page
2-6
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
rate, press Change
mL/h increments).
Enable
Disable
Change
Rate
Change
Volume
CONFIRM
volume, press Change
).
5%
Enable
Disable
Change
Rate
Change
Volume
CONFIRM
CONFIRM
displays.
1

2.4.6 Max rate

The Maximum rate value identifies the
maximum rate at which a bolus dose can be
delivered. When the Max Rate option is
selected during bolus dose programming,
to
0.01
the bolus is delivered at the maximum rate
for an automatically calculated duration.
1. After accessing System Config -
Syringe options display, page
Max rate soft key.
2. To change maximum rate, press
Change Value soft key and enter new
rate (
increments;
increments).
System Configuration - Syringe
Maximum rate
adjustment
Maximum
Change
rate:
Value
999
mL/h
Confirm
to
mL/h in
0.1
99.9
to
100
999
System Configuration - Syringe
Maximum rate
adjustment
Maximum
Change
rate:
Value
_ _ _
mL/h
-- Continued Next Page --
, press
2
mL/h
0.1
mL/h in