DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.8 Removing Logic Board Assembly

1. If present, remove Retainer Clip from
Dual Connector Cable-Logic Board
connection.
2. Disconnect Dual Connector Cable from
Logic Board
3. Disconnect Linear Sensor Harness from
Logic Board
4. Remove one (
Board to
IUI
5. Remove one (
Board to Internal Frame Assembly.
Internal Frame Assembly
Linear Sensor Harness
Dual Connector Cable
.
J8
.
J12
) screw attaching Logic
1
Support Bracket.
) screw attaching Logic
1
IUI Support Bracket
Retainer Clip
CORRECTIVE MAINTENANCE
6. Disconnect harnesses from following
locations on Logic Board:
(display)
J4
(motor)
J7
(flange detect)
J11
(lock harness)
J13
7. Remove Logic Board.
During Reassembly:
Route Dual Connector Cable and Linear
Sensor Harness under
J12
J8
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
PCA Module
Support Bracket.
IUI
Logic Board
J11
J4
J7
J13
5B-11