CORRECTIVE MAINTENANCE
PCA Module
DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.22 Removing Linear Sensor Assembly

CAUTION
To avoid damaging the Linear Sensor
Assembly, do not touch with bare hands.
1. Peel Linear Sensor Assembly off
Extrusion Frame.
2. Use alcohol to clean adhesive debris
from Extrusion Frame.
During Reassembly:
1. Ensure Extrusion Frame surface is
clean.
5B-26
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
Extrusion Frame
2. Before adhering Linear Sensor
Assembly to Extrusion Frame:
a. Remove protective backing from
Linear Sensor.
b. Align hole in Linear Sensor
Assembly with hole in Extrusion
Frame.
3. Use a burnishing tool to ensure that
Linear Sensor Assembly is securely
adhered to Extrusion Frame.
4. Clean Linear Sensor Assembly surface
using alcohol and lint-free cloth (such as
Kimwipes or lint-free tissue).
Linear Sensor Assembly
Alignment Hole