5.2.17 Removing Handset Harness

NOTE: Use a
least 4 in. deep to remove and install the
Handset Harness.
1. Ensure that
then remove top nut (first of
secures Handset Harness to Front
Case.
2. Use socket to remove second nut.
3. Remove Ground Harness.
4. Remove Handset Harness (already
disconnected from Display Board
/
-in. deep-well socket at
9
16
/
-in. socket is on top nut,
9
16
nuts) that
2
Nut
Nut
Ground Harness
During Reassembly:
1. Install Handset Harness with red dot
pointing to top of Front Case.
2. Install Ground Harness, then install first
nut and tighten.
3. Install second nut and tighten, ensuring
that socket is on top nut (second of
nuts) only.
4. Connect Handset Harness to Display
Board
).
J3