Priming; Configuration Setup - Shared Infusion Pump And Syringe; Configuration Options And Defaults - Pca Module - CareFusion Alaris 8110 series Service Manual

Syringe module; pca module
Hide thumbs Also See for Alaris 8110 series:
Table of Contents

Advertisement

2.4
CONFIGURATION SETUP - SYRINGE
MODULE (Continued)

2.4.10 Priming

Enabling Priming mode allows a limited
volume of fluid to be delivered to prime the
administration set before connecting to a
patient or following a syringe change. When
priming, a single continuous press of the
soft key delivers up to
PRIME
priming fluid.
1. After accessing System Config -
Syringe options display, page
Priming soft key.
System Config - Pump 2 of 3
System Configuration - Syringe
Priming mode
Allows use of the
priming feature.
2. To change option setting, press soft key
next to applicable option (Enable or
Disable).
• Option setting changes and System
Config - Syringe page
2.5
CONFIGURATION SETUP - SHARED
INFUSION - PUMP AND SYRINGE
MODULES
Refer to the
8000
Service Manual.
mL of
2
, press
2
Enable
Disable
displays.
2
/
Series Technical
8100
CHECKOUT AND CONFIGURATION
2.6
CONFIGURATION OPTIONS AND
DEFAULTS - PCA MODULE
NOTE: With the Profiles feature enabled, the
settings are configured independently for
each profile. A hospital-defined, best-
practice data set must be uploaded to enable
the Profiles feature. Date and Time is a
system setting and is the same in all profiles.
System Settings: Refer to the PC Unit
section of the Alaris
PCA Module Settings: Refer to the PCA
Module section of the Alaris
PCA Module Configuration Parameters:
Select system configuration settings for the
PCA Module during data set development,
then upload them as part of the data.
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
System
.
DFU
®
System
DFU
®
.
2-9