CORRECTIVE MAINTENANCE
Syringe Module
Level of Testing Guidelines - Syringe Module (Continued)
Table 5A-3.
Tests to Perform
= Required
X = If verification fails.
Blank = Not Applicable
Repair/Replacement of:
Miscellaneous:
Instrument Dropped
Instrument Software Flashed
New Instrument Checkout
No Fault Found
(instrument not opened)
No Fault Found
(instrument opened)
NOTES:
• Use the Maintenance Software to perform testing and calibration.
• After performing any calibration procedure, always perform the
associated verification test.