DISASSEMBLY/REASSEMBLY (Continued)
5.2

5.2.1 Removing Latch Assembly and Feet

1. Remove two (
Assembly to bottom of Rear Case and
remove Latch Assembly Components.
2. Pull two (2) Feet from underside of
module.
NOTE: When reinstalling the Compression
Spring, be sure to orient as shown during
reassembly.
) screws attaching Latch
2
Latch Screws
Compression Spring
Feet
CORRECTIVE MAINTENANCE
During Reassembly:
Apply thin layer of Dow Corning Molykote
(or equivalent) silicone grease to Feet.
NOTE: The Feet press-fit into the module.
Rear Case
Alaris
Syringe Module (8110 Series)
®
Alaris
PCA Module (8120 Series)
®
Technical Service Manual
Syringe Module
33
5A-3